打开APP
userphoto
未登录

开通VIP,畅享免费电子书等14项超值服

开通VIP
pch
4.6.1 Filtered output of Band 1-4 via an analogue AC output
With a PCH 1216 it is also possible to have the filtered output signal of any of the four (4) Bands
directly presented at an analogue AC output with a maximum measurable frequency of 1000 Hz.
The output signal of the SSD band is NOT available this way.
Please note that there are four bands but generally only two (optionally three) analogue outputs.
4.6.2 RAW data export via an analogue AC output
A PCH 1216 or PCH 1218 has the possibility to export RAW data that is taken from the device
after the RAW band filter. The maximum measurable frequency is 35 Hz. However, unless the
optional 3rd analogue DC output is mounted and enabled, it is only possible to receive RAW data
from any two of the three directions.
PCH 1216
Page 18
4.7 CANopen (pin #5+6+8 or #6+7+8))
The PCH 1218 device (only) can communicate with a PC or PLC by using the CANopen protocol.
The use of the CANopen feature requires three connections to be made:
- CANopen – LOW at
~ pin #5 for a PCH 1218 with an 8-pin connector
~ pin #7 for a PCH 1218 with a 12-pin connector
- CANopen – HIGH at
~ pin #8 for a PCH 1218 with an 8-pin connector
~ pin #8 for a PCH 1218 with a 12-pin connector
- CANopen – GND at
~ pin #6 for a PCH 1218 with an 8-pin connector
~ pin #6 for a PCH 1218 with a 12-pin connector
See sections 4.1 and 4.2 for the layout of the connectors that are relevant for your PCH 1216 or
PCH 1218 device.
There is one receive PDO and there are nine transit PDOs with the following definitions:
- RECEIVE PDO : Alarm inhibit band 1-4; Ramp-test; SSD-test; Self-test; Alarm reset
- TRANSIT PDO1 : Raw data from sensor 1, one data sample every 10 ms
- TRANSIT PDO2 : Raw data from sensor 2, one data sample every 10 ms
- TRANSIT PDO3 : Raw data from sensor 3, one data sample every 10 ms
- TRANSIT PDO4 : Internal Compatibility Word (ICW); Optimized status 1 (= monitor status);
Optimized status 2 (= general status band 1-4 plus SSD band); Danger alarm
status 1-4 plus SSD; Alert alarm status 1-4
- TRANSIT PDO5 : Band SSD measuring result; measurement unit; sequence number
- TRANSIT PDO6 : Band 1 measuring result; measurement unit; sequence number
- TRANSIT PDO7 : Band 2 measuring result; measurement unit; sequence number
- TRANSIT PDO8 : Band 3 measuring result; measurement unit; sequence number
- TRANSIT PDO9 : Band 4 measuring result; measurement unit; sequence number
By setting/toggling certain bits in the PDO using software CHT 5017 (CANopen master control and
display program), which has been released by the manufacturer PCH Engineering A/S, it is possible
to inhibit the alarms of all four Bands 1-4 individually, carry out a ramp-test, initiate a Self-Test,
initiate a general Alarm re-set (NOT SSD Danger alarm reset) or initiate a dedicated SSD-Test (i.e.
also SSD re-set).
!
Warning:
In principal – according to the GL2010 guide lines – the SSD test/re-set is to be carried
out by a technician that is personally present in the wind turbine (nacelle). However, if
the wind turbine is densely surveyed by means of cameras (and microphones) that are
positioned both inside and outside the wind turbine thus giving a technician in a remote
control room a full overview of the wind turbine’s operational condition, it is allowed to
initiate the SSD test from this remote control room. See the GL2010 certification guide
lines, chapter 2.2.2.5 on the subject of camera surveillance.
Page 19
!
Note:
By default the PCH 1218 does NOT come with an internal 120 Ω termination resistor
installed. However, such a resistor must be installed in a PCH 1218 device that happens
to be the last device in the CanBus chain. When ordering please specify the presence of a
120 Ω termination resistor by adding R0 (not installed, default) or R1 (installed) to the
CHF838xxx number: CHF838xxxR0 or CHF838xxxR1. See your Customer Specification Agreement (section 2) for more information on this topic.
4.8 Power/Digital/Signal GND (pin #7 or #10)
Digital input signals and analogue DC outputs are all relative to the ground potential on:
- pin #7 for PCH 1216 and PCH 1218 devices with an 8-pin connector
- pin #12 for PCH 1216 and PCH 1218 devices with a 12-pin connector
Pin #7 or pin #12 is also the power ground potential for the +24 VDC connection at pin #1.
See sections 4.1 and 4.2 for the layout of the connectors that are relevant for your PCH 1216 or
PCH 1218 device.
4.9 Self-Test (pin #8)
The PCH 1216 (only) is provided with a pin in its connector that is dedicated to start a Self-Test of
the device by connecting this pin with digital GND. The Self-Test function tests the integrity of the
total device, incl. the transducers:
- pin #8 for all PCH 1216 devices
The corresponding digital ground (GND) is available at:
- pin #7 for a PCH 1216 device with an 8-pin connector
- pin #12 for a PCH 1216 device with a 12-pin connector
See sections 4.1 and 4.2 for the layout of the connectors that are relevant for your PCH 1216 or
PCH 1218 device.
If the test is successful the unit will – after a short settling time – return to the stand-by mode,
indicated by a green status LED on the enclosure.
If the test is not successful the unit will – after a short settling time – go to the System Error mode,
indicated by a red status LED on the enclosure.
Only individual functions/features can be addressed during a self-test. But there are NO
consequences, i.e. no alarms will be triggered if a certain vibration level set during self-test exceeds
an alarm trigger level. If an alarm must be triggered during self-test the alarm in question must be
marked. The behaviour of outputs during a self-test must also be specified separately.
4.9.1 Self-Test using the optional RS485 ModBus connection
Both a PCH 1216 and a PCH 1218 can be set into self-test mode using a mounted (optional) and
enabled RS485 ModBus connection. The self-test can then e.g. be started by the PCH Vibration
Studio software.
4.9.2 Self-Test using the CANopen FieldBus protocol
The RECEIVE PDO includes a bit that is used for initiating and then stopping a self-test of a PCH
1218 device (only). See section 4.7 of this manual.
Page 20
4.10 RS485 ModBus communication (pin #5 and #9)
Optionally any PCH 1216 or PCH 1218 can be purchased with a RS485 ModBus function for
communication with a PC or a PLC. Such a device will always have a 12-pin connector.
The RS485 ModBus function is a useful feature in the R&D phase of a new model of wind turbine
because it allows full control of the device by PC using the PCH Vibration Studio software, which
has been released by the manufacturer PCH Engineering A/S.
The PCH Vibration Studio software in connection with a RS485 Modbus function has the
possibility to read measuring results, read the status flags of many functions/features in the device
and last but not least allows the operator to alter the parameter set-up of the device in order to
optimize it for the wind turbine (machine).
To connect a PCH 1216 or PCH 1218 device to the RS485 field bus requires two connections:
- RS485 +: pin #5 for a PCH 1216 or PCH 1218 device with a 12-pin connector
- RS485 –: pin #9 for a PCH 1216 or PCH 1218 device with a 12-pin connector
See sections 4.1 and 4.2 for the layout of the connectors that are relevant for your PCH 1216 or
PCH 1218 device.
!
Note:
The RS485 ModBus maybe used while all measuring chains are operational.
!
Note!
By default a PCH 1216 or PCH 1218 does NOT come with an internal 120 Ω
termination resistor installed. However, such a resistor must be installed in case the
device happens to be the last device in the RS485 chain. When ordering please specify
the presence of a 120 Ω termination resistor by adding R0 (not installed, default) or R1
(installed) to the CHF836xxx or CHF838xxx number, e.g CHF836xxxR0 and
CHF838xxxR1. See your Customer Specification Agreement (section 2) for more
information on this topic especially if there is a RS485 ModBus and a CANopen
FieldBus installed in the same device.
4.11 Light Emitting Diode (Status LED) on enclosure
On the stainless steel enclosure, just above the connector, you will find a LED that is used to
indicate the current status of a PCH 1216 or PCH 1218:
Status
: Green:
- Static
: Normal operation. Power ON
- Flashing
: Settling
Red:
- Static
: SSD alarm or System Failure
- Flashing
: Update mode
Yellow:
- Static
: SSD Danger alarm relay test
- Flashing
: Sensor test
Page 21
4.12 CHB 1131 Service Cable
The CHB 1131 Service Cable allows easy direct communication between the USB 2.0 port of a PC
and one PCH 1216 or PCH 1218 device . A CHB 1131 will also provide main power to the device.
There are two versions of the Service Cable available:
- CHB 1131P8: USB 2.0 Service Cable for a PCH 1216 or PCH 1218 with a 8-pin connector
- CHB 1131P12: USB 2.0 Service Cable for a PCH 1216 or PCH 1218 with a 12-pin connector
The CHB 1131 has a small push button that when pressed shortly will start a SSD self-test.
Together with the PCH Vibration Studio software the CHB 1131 Service Cable can be used to gain
full control over a PCH 1216 or PCH 1218 but….. only one at the time.
The CHB 1131 Service Cable also enables (individual) communication with PCH 1216 or PCH
1218 devices that elsewhere do not have a communication feature included in their hardware and/or
firmware.
!
Note!
Please be advised that when using a CHB 1131 Service Cable, the connected device will
be in off-line mode.
4.13 CHB 1134 Service Adaptor
The CHB 1134 Service Adaptor allows easy direct communication between the USB 2.0 port of a
PC and one PCH 1216 or PCH 1218. A CHB 1134 will also provide main power to the device.
Page
5 System Failure Monitoring
5.1 System Failure Monitoring
System Failure Monitoring is a self-monitoring so-called watchdog feature of both the PCH 1216
and the PCH 1218 devices. It ensures that under no circumstance can an erroneous measurement
occur without causing a System Failure.
The continuous operative self-monitoring System Failure procedure is based on a number of
possible error conditions, as described in the sections below:
5.1.1 Errors related to the set-up of the monitor and the input signal
The System Failure system detects that the measurement (results) could be compromised.
- Overload in input section (accelerometer signal too large)
- Overload of DC output
Remedy: Both errors can be corrected by checking the cables and/or the set-up of the monitor.
5.1.2 Errors related to the environmental conditions
The system Failure system detects that something is wrong with the external conditions under
which the device operates:
- Power failure
- Internal temperature exceeding limits
Remedy: Both errors can be corrected by improving the environmental conditions.
5.1.3 Errors related to the hardware and the firmware of the Monitor
The System Failure system detects a malfunction of the PCH 1216 or PCH 1218.
- Error in set-up data
Remedy: Reload the set-up. If this does NOT help, the unit must be returned to the manufacturer
for repair. Please refer to section Error! Reference source not found. of this User
anual for more details on how to get your device serviced/repaired.
5.1.4 Errors that only PCH Engineering A/S service personnel can correct
- Persistent error detected by Self-test
- Error in factory calibration data
- Digital Signal Processor stop or error in hardware and/or firmware
Remedy: Only possible at factory level. Please return the PCH1216/1218 to the manufacturer for
repair. Please refer to section Error! Reference source not found. of this User Manual
or more details on how to get your device serviced/repaired.
216/1218 structural vibration monitor with SSD function
Page 24
5.2 System Failure Relay
The PCH 1216 and PCH 1218 devices do NOT have a dedicated System Failure relay. Instead a
PCH 1216 or PCH 1218 device uses the SSD Danger relay(s) for this purpose. The System Failure
(= SSD) relay(s) trigger on any of the system failure conditions described above. The System
Failure Relay remains active until the reason for the fault is eliminated and the same relay(s) are
therefore NOT latched like when they are activated by a SSD alarm.
The System Failure Relay is energized (i.e. in make condition) when no system failures are active
(stand-by), and de-energized (i.e. in break condition) when in system failure mode (activated). In
this way the relay will also indicate a system failure when the power goes off, which is a system
error by itself.
!
Warning!
The relay(s) are so-called contact relay(s). Excessive current (and/or voltage) will
destroy the relay. For this reason please do not apply any power source directly to the
relay. Protect the relay with a resistor in series that will limit the current.
Please make sure that the following ratings/conditions for the relay(s) are not exceeded in your
application:
- Max load voltage : 48 VDC
- Maximum current load : 100 mA
- Maximum power load : 5 Watts
5.3 System Failure Indicator
The status LED on the enclosure will turn red as soon as a System Failure occurs. At the same time
“System Failure” indicator(s) within the PCH Vibration Studio software will turn red.
5.4 Finding the reason of a System Failure
Detailed information concerning the reason that might have caused the System Failure alarm can be
found at the bottom of the “Results” display that belongs to the PCH 1216 or PCH1218 in question
in the PCH Vibration Studio software:
Another source of useful information could be available in the columns “Category”, “Description”
and “Source” of the “Status and Events” display in the same software:
Page 25
6 Power Source
6.1 Voltage Range
A PCH 1216 or PCH 1218 must be powered by a +24 VDC (i.e. 10,5 – 30,0 VDC) power supply
via its pin #1.
Power GND can be found at:
- pin #7 for a PCH 1216 or PCH 1218 device with an 8-pin connector
- pin #12 for a PCH 1216 or PCH 1218 device with a 12-pin connector
See sections 4.1 and 4.2 for the layout of the connectors that are relevant for your PCH 1216 or
PCH 1218 device.
For more details on this subject, please read sections 4.3 (on +24 VDC) and 4.8 (on GND) of this
manual.
6.2 Power ON LED
When the status LED on the enclosure of a PCH 1216 or PCH 1218 device is radiating green,
yellow or red light – either continuously or in flash mode – this indicates that the device is properly
powered.
Page 26
7 Connecting for (remote) control
A PCH 1216 or PCH 1218 can be remotely controlled:
- when part of a RS485 ModBus chain of devices. Such a data connection is first-off-all intended
for reading out result data, as well as status information, but can also be used for set-up and
(remote) field maintenance. On-line communication is possible while the device is operational
- using the CANopen FieldBus protocol. This way of communication is limited to collect raw
vibration level data directly from the sensor, measuring results from the four conditioning bands,
as well as status concerning the SSD Danger alarm. In addition it is possible to start a Self-Test,
a SSD-test or a Ramp-test. The CANopen protocol does not allow changing the parameter setup
of the device!!! On-line communication is possible while the device is operational
- using a LAN connection with an optional CHB 1120 Etherbridge. The CHB 1120 will be
connected on one side to the on-site LAN while the other side is connected to either one (max
two) RS485 ModBus chains of PCH 1216 and/or PCH 1218 device(s). Using this way of
communication gives full control of one or many PCH 1216 and/or PCH 1218 devices at the
same time. On-line communication is possible while the device is operational
- using a USB connection with an optional CHB 1120 Etherbridge. The CHB 1120 will be
connected on one side to the USB 2.0 connector of a (tablet-)PC while the other side is
connected to either one (max two) RS485 ModBus chains of PCH 1216 and/or PCH 1218
device(s). Using this way of communication gives full control of one or many PCH 1216 and/or
PCH 1218 devices at the same time. On-line communication is possible while the device is
operational
- using the correct version of CHB 1131 Service Cable or CHB 1134 Service Adaptor in off-line
mode!!! See section XXX for more information
!
Warning!
It is the operator’s responsibility to use a PC that is conforming to the standard EN/IEC
60950 and that the environment in which the PC is located is within the temperatureand humidity range specified by the PC manufacturer. Also make sure that the PC is
still conform standard EN/IEC 60950 when connected to the PCH 1026 Mk2.
All input/output registers can be accessed from the main ModBus interface, but only a subset
(SubBus) can be accessed from the MultiBus (Fieldbus) interface
CHO 1115-01
Measurement
Setup
MultiBus
CANopen or Interbus,
Profibus or CAN,
or Devicenet or ....
Measurement
Control
MultiBus
Setup
SubBus
RAW AC 10ms
Direct from DSP
ModBus
PCH 1026
ModBus
RS232 or
RS485
FieldBus
Page 27
PC PCH1216/18
Terminal DATA +
Terminal DATA -
120 W Termination resistors
Master Slave(s)
7.1 Data transfer via RS485 Modbus RTU (Optional)
A PCH 1216 or PCH 1218 device is capable of remote communication via an optional RS485 Serial
Multi-drop interface at distances of up to 1200 m. Connection of more than 32 slaves, or wiring
over distances of more than 1200 m, is only possible in combination with a RS485 repeater. Up to a
maximum of 247 slaves can be addressed in one RS485 chain.
Always connect the RS485 ModBus cable to:
- RS485+ at pin #5 for a PCH 1216 or PCH 1218 device with a 12-pin connector
- RS485– at pin #9 for a PCH 1216 or PCH 1218 device with a 12-pin connector
See sections 4.1 and 4.2 for the layout of the connectors that are relevant for your PCH 1216 or
PCH 1218 device.
!
Note!
For your RS485 connections, please always use a RS485 certified two-wire twisted and
shielded cable.
To ensure the best noise immunity, the cable has to be terminated with a 120 W resistor at both ends
of the chain, i.e. immediately after the connection to the PC/PLC and immediately before the
connection to the last device in the chain.
!
Note!
By default a PCH 1216 or PCH 1218 does NOT come with an internal 120 Ω
termination resistor installed. However, such a resistor must be installed in case the
device happens to be the last device in the RS485 chain. When ordering please specify
the presence of a 120 Ω termination resistor by adding R0 (not installed, default) or R1
(installed) to the CHF836xxx or CHF838xxx number, e.g CHF836xxxR0 and
CHF838xxxR1. See your Customer Specification Agreement (section 2) for more
information on this topic especially if there is a RS485 ModBus and a CANopen
FieldBus installed in the same device.
PCH1216/18
Connection to RS485 ModBus
Page 28
All registers can be accessed via the PCH1216/1218 ModBus interface. The RS485 interface
includes a register address range referred to as SubBus. Registers in the SubBus area are also
available on the MultiBus interface.
The RS485 ModBus address can be set from 1 to 247. (Default = 1)
The baud rate can be set between 300 and 38400 baud (Default = 38400)
!
Warning!
Readable data will only be present at the SubBus if the PCH1216 or PCH 1218 device is
in measuring mode. This prevents false readings.
!
Warning:
Please make sure that you do not add a new device to your FieldBus that has an address
occupied by another device already present in the RS-485 chain of devices.
7.2 Data transfer via CANopen
The CANopen standard used is: CIA DS301 revision 4.
All the data from the SubBus can be transmitted via the CANopen interface, including
measurement results, alarms and status. Also a 10ms sample rated RAW AC signal for each of the
three vibration directions A, B and C is available on the CANopen.
To run the monitor in a CANopen network, it is necessary to set the Node Address so it will not
conflict with other components on the same segment of the bus. This is easily done by PCH
Vibration Studio software with ModBus access via RS485 or by using a CHB 1131 Service Cable
or CHB 1134 Service Adaptor. The default node address is: 2.
7.2.1 CANopen MultiBus protocol
A PCH 1218 is equipped with a set of registers called SubBus registers. These registers are meant
as an easy and fast option to get essential status and selected results from the monitor.
It is possible to configure the MultiBus interface via the RS485 ModBus interface from a PC/PLC
or other similar equipment, but the easiest way to set up everything around the SubBus/MultiBus is
with the PCH Vibration Studio software.
!
Warning!
Values will only be present at the SubBus if the PCH 1218 has been or is in Measuring
State. This prevents false readings.
!
Warning!
Only results defined in the SubBus area are available at the MultiBus interface (the
installed Fieldbus).
Whether or not a particular PCH 1218 device has a RS485 ModBus installed can be seen from type
number label, which indicates the presence of the termination resistor. Another good (but NOT
conclusive) indicator of an embedded RS485 ModBus RTU in a PCH 1218 device is the presence
of a male 12-pin connector on the side panel of the enclosure.
Page 29
7.2.2 Input/Output data transfer
In case you need more information than described below on how to make a PCH 1218 device
communicate with a PC or PLC using the CANopen protocol, please contact the manufacturer PCH
Engineering A/S for more detailed information.
!
Note:
By default the PCH 1218 does NOT come with an internal 120 Ω termination resistor
installed. However, such a resistor must be installed in a PCH 1218 device that happens
to be the last device in the CanBus chain. When ordering please specify the presence of a
120 Ω termination resistor by adding R0 (not installed, default) or R1 (installed) to the
CHF838xxx number: CHF838xxxR0 or CHF838xxxR1. See your Customer Specification Agreement (section 2) for more information on this topic.
There is only one RECEIVE PDO and but there are nine TRANSIT PDOs.
By setting/toggling certain bits in the SEND PDO using software CHT 5017 (CANopen master
control and display program), which has been released by the manufacturer PCH Engineering A/S,
it is possible to inhibit the alarms of all four Bands 1-4 individually, carry out a ramp-test, initiate a
Self-Test, initiate a general Alarm re-set (NOT SSD Danger alarm reset) or initiate a dedicated
SSD-Test (incl. a SSD alarm re-set).
- RECEIVE PDO : Alarm inhibit band 1-4; Ramp-test; SSD-test; Self-test; Alarm reset
- TRANSIT PDO1 : Raw data from sensor 1, one data sample every 10 ms
- TRANSIT PDO2 : Raw data from sensor 2, one data sample every 10 ms
- TRANSIT PDO3 : Raw data from sensor 3, one data sample every 10 ms
- TRANSIT PDO4 : Internal Compatibility Word (ICW); Optimized status 1 (= monitor status);
Optimized status 2 (= general status band 1-4 plus SSD band); Danger alarm
status 1-4 plus SSD; Alert alarm status 1-4
- TRANSIT PDO5 : Band SSD measuring result; measurement unit; sequence number
- TRANSIT PDO6 : Band 1 measuring result; measurement unit; sequence number
- TRANSIT PDO7 : Band 2 measuring result; measurement unit; sequence number
- TRANSIT PDO8 : Band 3 measuring result; measurement unit; sequence number
- TRANSIT PDO9 : Band 4 measuring result; measurement unit; sequence number
7.2.3 Export of RAW data from sensors
A PCH 1218 device has the possibility via TRANSIT PDO1, PDO2 and PDO3 to export RAW data
that is taken from the device after the filter but before the detector for each of the three (3) sensors
(A, B and C) respectively.
The data is presented in 10 ms intervals, i.e. 100 times per second. According to the Nyquist
theorem this would allow the transfer of a maximum signal frequency bandwidth of 50 Hz, but 35
Hz is a more practical value. The maximum measurable frequency is therefore limited to 35 Hz.
Page 30
7.2.4 Status and Alarms
RECEIVE PDO4 contains the following information that is available on the SubBus:
- Internal Compatibility Word (ICW)
- Optimized status 1 (i.e. monitor status)
- Optimized status 2 (i.e. general status in conditioning band 1-4 plus SSD band)
- Danger alarm status in conditioning bands 1-4 plus SSD band
- Alert alarm status in conditioning bands 1-4
Outside the SubBus range there will be access to more detailed status. The by default implemented
status information’s should be sufficient in most cases. However, in case the operator needs to read
out not (yet) implemented status information, please consult the manufacturer PCH Engineering
about a possible implementation.
7.2.5 Measurement results
RECEIVE PDO5 through PDO9 contains the measurements of the SSD band and the four (4)
conditioning bands respectively. Also the measurement unit is sent each time, as well as sequence
number that was increased by one (1) after each completed cycle of nine PDOs since the data
collection started.
7.2.6 Self-Test
It is possible to bring the PCH 1218 device in self-test mode by setting the correct bit in the
RECEIVE PDO. Toggling this bit will stop the self-test mode.
7.2.7 SSD-Test and SSD reset
It is possible to bring the PCH 1218 device in SSD-test mode by setting the correct bit in the SEND
PDO. Toggling this bit will stop the self-test mode.
A successfully ended SSD-Test is the only method to make an end to (i.e. reset) an activated SSD
alarm.
!
Warning:
In principal – according to the GL2010 guide lines – the SSD test/re-set is to be carried
out by a technician that is personally present in the wind turbine (nacelle). However, if
the wind turbine is densely surveyed by means of cameras (and microphones) that are
positioned both inside and outside the wind turbine thus giving a technician in a remote
control room a full overview of the wind turbine’s operational condition, it is allowed to
initiate the SSD test from this remote control room. See the GL2010 certification guide
lines, chapter 2.2.2.5 on the subject of camera surveillance.
7.2.8 Communications settings
The Node Address can be set to a value between 1 and 126. (default: 2)
The Baud rate must also be set, 0..8 ~ [10k,,1000] kBaud. (default 500 kBaud)
Page 31
!
Warning!
It is not recommended to choose a Baud rate below 125 KBaud with all PDO's 1-9
activated.
!
Warning!
After changing the Baud rate the vibration monitor must be restarted (e.g. by carrying
out a power OFF/ON).
!
Note!
Setting of a new node address or new Baud rate is only possible using a CHB1131
Service Cable, a CHB 1134 Service Adaptor (off-line) or a PCH 1218 with a mounted
and enabled RS-485 ModBus interface (on-line).
7.2.9 NO changing of the parameter setup
A basic PCH 1218 device has NO communication facility to alter the parameter setup. For this a
CHB1131 Service Cable, a CHB 1134 Service Adaptor or a PCH 1218 with a mounted and enabled
RS-485 ModBus interface will be required.
!
Warning!
While changes in the parameter setup using a CHB 1131 or CHB 1134 will be off-line,
they might be carried out on-line with the RS-485 interface, i.e. while all measuring
chains are operational.
7.3 Data transfer via LAN
Any PCH 1216 or PCH 1218 device that has an embedded RS485 ModBus feature can be remotely
controlled/operated using a LAN connection. However, this requires the use of an additional
optional CHB 1120 PCH EtherBridge module.
All PCH 1216 or PCH 1218 devices with embedded RS485 ModBus will have a 12-pin connector.
Page 32
The PCH EtherBridge module is on one side connected to a Local Area Network (LAN) with a
LAN cable. The other side of the EtherBridge has 2 COM ports that both can be configured to act
as the main control unit in a separate RS485 daisy chain of ModBus devices.
On the CHB 1120 EtherBridge side the RS485 port is connected as follows:
For a PCH 1216 or PCH 1218:
- COM1 – pin #14: RS485+ of Modbus chain 1; PCH 1216 or PCH 1218 device pin #5
- COM1 – pin #15: RS485– of ModBus chain 1; PCH 1216 or PCH 1218 device pin #9
- COM1 – pin #16: GND (cable shield) of ModBus chain 1
- COM2 – pin #17: GND (cable shield) of ModBus chain 2
- COM2 – pin #18: RS485+ of Modbus chain 2; PCH 1216 or PCH 1218 device pin #5
- COM2 – pin #19: RS485– of ModBus chain 2; PCH 1216 or PCH 1218 device pin #9
See sections 4.1 and 4.2 for the layout of the connectors that are relevant for your PCH 1216 or
PCH 1218 device.
It is allowed to connect a PCH 1216 device to COM1 and a PCH 1218 device to COM2 and v.v.
!
Warning!
To increase data transfer reliability please use a two wire twisted and shielded RS485
certified cable.
!
Warning!
To reduce (i.e. improve) the noise sensitivity please do not forget to terminate both
ends of your RS485 ModBus chain with a 120 Ω resistor. Both COM ports of a CHB
1120 have an embedded 120 Ω termination resistor that can be enabled by visiting the
“homepage” available directly in a browser at the IP address of the device or via the
PCH Vibration Studio software while using a LAN connection.
For more information on how to place a PCH 1216 or PCH 1218 device in a chain of devices
connected through a RS485 ModBus, please read section 7.1 of this manual.
For more detailed information concerning the operation of the CHB 1120 PCH EtherBridge we
recommend to read the CHB 1120 PCH EtherBridge user manual.
7.4 Data transfer via CHB 1131/1134 Service Cable/Adaptor
By using an optional Service Cable or Service Adaptor to establish a direct connection between an
USB 2.0 port of a notebook/tablet PC and the circular 8-pin or 12-pin connector on the enclosure of
a PCH 1216 or PCH 1218 device, it is possible to obtain two-way communication with this device.
Whenever a PCH 1216 or PCH 1218 device is connected to a PC by the Service Cable or Service
Adaptor the device will also get its +24 VDC main power this way.
PCH Vibration Studio software facilitates the collection of measurement results and status read-out.
The same software can also be used to make changes in the parameter set-up of a PCH 1216 or
PCH 1218 device.
There are two different Service Cables available:
- CHB 1131P8 : For PCH 1216 or PCH 1218 devices with an 8-pin connector
- CHB 1131P12 : For PCH 1216 or PCH 1218 devices with a 12-pin connector
Page 33
There are also two different Service Adaptors available:
- CHB 1134P8 : For PCH 1216 or PCH 1218 devices with an 8-pin connector
- CHB 1134P12 : For PCH 1216 or PCH 1218 devices with a 12-pin connector
The 8-pin version can easily be changed into a 12-pin version of the Service Adaptor by
The Service cable becomes very useful if frequent parameter changes are required, e.g. in the
development phase of a new wind turbine or for a quick update of a PCH 1216 or PCH 1218 device
in the field in connection with the change of one or more set-up parameter(s) or an update/upgrade
of the firmware.
Page 34
8 PCH 1216/1218 and PCH Vibration Studio
The PCH Vibration Studio software is delivered on a DVD. The installation of the software is
straight forward and described in detail in the ”CHF 2042 – User Manual for CHT 1044 PCH
Vibration Studio User Software for PCH Vibration Monitors”.
The PCH Vibration Studio software is available in two versions: a Standard version and an OPC
version.
!
Note!
Any PCH 1216 or PCH 1218 can only be operated together with the OPC version of the
PCH Vibration Studio software.
The installation process automatically installs all necessary drivers.
In order to help the customer maintain the original parameters set-up as unchanged as possible, a
read-only version of the PCH Vibration Studio software is available, where set-up parameters
cannot be changed but only read-out. This is intended for those service personnel who are not
authorized to change parameters. All of our customers are quite happy with this set-up, making
them confident that quality is managed well.
Page 35
The PCH Vibration Studio software is a powerful tool to:
- Get a direct readout of the measured vibration parameter(s) of any PCH Engineering vibration
monitor device, including that of PCH 1216 and PCH 1218 devices
- Follow in-time the trends of vibration parameters, incl. alarm indication
- Machine vibration diagnosis by analysing the vibration results of groups of transducers
- Readout the status of the device making it possible to see the reason for certain error indications
like input signal overload and power cut
- Forced (remote) self-test of the device and see how the device reacts under stress
- Change of setup parameters in the laboratory or in the field. Please note: that such action voids
the Customer Specification Agreement setup, if any.
!
Note!
Please read the PCH Vibration Studio user manual on how to install this software on
your (tablet-)PC. The sections below only describe the basics with regard to how to
parameterise a PCH 1216 or PCH 1218 device. For more detailed information or
subjects like Trend readings, data streaming, off-line FFT analysis, etc. also please
consult the PCH Vibration Studio software User Manual.
All screen dumps shown in this section of the manual are taken from a PC running the PCH
Vibration Studio software.
8.1 General
It is possible to change one or more setup parameters after taken delivery of a PCH 1216 or PCH
1218 device. For this purpose, please run the PCH Vibration Studio software, which must be
installed on a PC that is connected to an on-site PCH 1216 or PCH 1218 device via its optional
embedded RS485 ModBus interface, via an optional LAN connection using a CHB 1120
EtherBridge, via an optional CHB 1131 Service Cable or an optional CHB 1134 Service Adaptor.
Although many PCH Engineering devices might be visible in the PCH Vibration Studio software
simultaneously, any change of setup parameters can only be carried out for one device at the time.
On the other hand measurement results and status data can be collected sequentially from all PCH
Engineering vibration monitors that are listed under “Devices”.
Some features/functions described in this manual might not be available or have been disabled by
the manufacturer in the device delivered to you prior to delivery. Only those measuring parameters
concerning features/functions that have been installed and enabled can be changed. Depending on
the available hardware, certain features/functions can be implemented in the future, but only at the
location of the manufacturer as a chargeable upgrade. The description of the measuring parameter
setup in this chapter is based on using the PCH Vibration Studio software for this purpose.
If the value of an entered parameter is invalid the field will turn Red and if the mouse cursor is
moved to the red coloured full scale field, a range of valid values will pop up.
!
Note!
It is strongly recommended to check the integrity of a parameter setup by clicking on
the “Validate!” button before applying the settings to your PCH 1216 or PCH 1218
device.
Page 36
8.2 Monitor
In the PCH Vibration Studio software “Edit/Monitor” page it is possible to see those
features/functions that are available for use.
Fields that are marked indicate features that are mounted.
Some features that are not mounted have missing hardware components on the main printed circuit
board of the device. This type of features can never – even by the manufacturer – be enabled after
delivery has taken place.
Alarm Hang Time
The “Alarm Hang Time” is the time an alert or danger alarm remains activated even though the
vibration level has decreased to below the trigger level. The hang time gives a PLC time to react on
short lived alarms. The alarm hang time is set for all alarms at once at a typical default value of: 1
second.
However, the system failure relay (and LED indication) has NO hang time and will remain activated
until the system failure condition has ended.
Overload Hang Time
The “Overload Hang Time” is the time a sensor-overload indication remains activated even though
the overload condition has ended. The hang time gives a PLC time to react on short lived sensoroverload indications. The alarm hang time is set for all sensor overloads at once at a typical default
value of: 1 second.
Page 37
8.3 Self-Test
Both the PCH 1216 and PCH 1218 include a “Self-Test” function that can be activated by using –
remote or on-site – the PCH Vibration Studio software with either a RS485, LAN or Service
Cable/Adaptor connection. For more information about Self-Test, please read section 4.9 of this
user manual.
The user/operator will have to specify/set the following in connection with a self-test:
- A certain pre-set scalar value during self-test of the RAW SampleValue for each of the three
(3) pairs of accelerometer transducers
- Whether or not the sensor-overload indicator related to each of the six (6) accelerometer
transducer must be activated during self-test
- The output level of the DC analogue outputs during self-test as a percentage of full scale of DC
current (mA) or DC voltage (V)
- Whether or not the sensor-overload indicator related to each of the two (2) analogue outputs
must be activated during self-test
- A certain pre-set measurement result scalar level for each conditioning band in its own units,
e.g. mm/s2, mm/s or mm
- The state of the Alert and Danger alarms during the self-test (note: During self-test the alarm states
must be specified separately – they are not automatically generated by any preset measurement
results level exceeding an alarm limit)
Please find on the below a complete overview of the values/parameters related to the self-test
procedure of a PCH 1216 or PCH 1218.
Page 38
The “Status” LED on a PCH 1216 or PCH 1218 device will will flash Yellow during a self-test.
! Warning:
Before setting the self-test parameters it is strongly recommended to make sure that you
understand the meaning of each parameter and the consequences the value and/or
status of certain parameter(s) will have on your complete installation.
8.4 Setup of general conditioning Bands 1 - 4
A PCH 1216 or PCH 1218 device has five (5) conditioning bands of which four (4) are for general
vibration monitoring use.
The fifth conditioning band is dedicated to the SSD Danger alarm function (only). See section 9.
Transducer:
Choose the transducer(s), i.e. vibration directions, to be monitored. There is a choice of any of the
three vibration vectors A, B or C (SingleDir mode), the sum of any two vibration vectors in one
plane A+B, B+C or A+C (OmniDir mode) or the sum of all three vibration vectors A+B+C
(TriaxDir mode).
The choice “None” will disable the conditioning Band.
Whenever one of the four general conditioning Bands is disabled, the respective vibration
parameter will disappear from the main Result display of the PCH Vibration Studio Software.
The conditioning Band reserved for the SSD Danger alarm function cannot be disabled.
The conditioning band reserved for the SSD function is default set to TriaxDir mode. Any of the
OmniDir modes is also allowed, but SingleDir modes are not part of the SSD function.
Page 39
Name:
It is recommended to give each conditioning Band a useful and easy recognizable name.
Measurement Setup
Unit:
Choose a measuring unit for each conditioning Band:
- Acceleration : m/s2,mm/s2, mg or g
- Velocity
: mm/s
- Displacement : mm or μm
Full Scale:
Enter the full scale of your choice. However, there are certain limitations related to the
accelerometer transducer. In case you make an erroneous choice the field will turn red and if your
curser is held over the field you will get information on valid scalar values that can be entered.
Filters:
When clicking on the Filter Range field the following should pop-up:
Before defining the band filter of the conditioning Band it is necessary to choose from two (2) sets
of filter ranges. The “Showing 0.1 to 120 Hz filters” is a set of filters used for wind turbine
structural vibration monitoring. The second “Showing 0.4 to 1 kHz filters” is used to monitor the
vibration of machinery, e.g. the gearbox, generator or bearings.
Now make your choice of band-pass filter within the range just defined. This is done by first
marking a Low Limit Filter (HP-High Pass) on the left side and then an Upper Limit Filter (LP-
Page 40
Low Pass) on the right side. After you have made your choice you can leave this frame by clicking
on the “Close” button.
Once you have made your choice of band-pass filter it is possible to obtain a detailed description of
this band-pass filter by clicking on the “Filter Info” button.
Detector:
The four (4) general conditioning bands in a PCH 1216 or PCH 1218 have a choice from four (4)
different detector types:
In general only the detectors RMS, Peak and Peak-to-Peak (P-P) are used. (Default: RMS with an
averaging time of 3 s).
The fourth detector is an (optional) RMS detector with 600 seconds averaging time, which is used
for very special applications only.
Page 41
The decision with regard to which detector to choose in each of the four general conditioning Bands
in a PCH 1216 or PCH 1218 device depends on the application of which the device will be a part:
- True RMS detector:
The most common detector is the True RMS detector.
By default the detector for any general conditioning Band will be pre-set to “RMS”
The only additional parameter that will have to be set is the RMS Averaging Time (in seconds).
By default the value of the RMS Averaging Time is pre-set to: 3 seconds.
- Peak detector:
The Peak detector has two additional parameters that will have to be set:
~ Attack Time (in milliseconds):
If nothing is specified the Attack Time will by default be pre-set to 5.0 milliseconds
~ Decay Time (in seconds):
If nothing is specified the Decay Time will by default be pre-set to 2 seconds
- Peak-to-Peak detector:
The Peak-to-Peak detector is setup in exactly the same way as the Peak detector.
Alert Alarm Setup:
It is possible to attach an Alert alarm to the above mentioned detectors that will become active as
soon as a pre-set trigger level has been exceeded for a period of time that is longer than the Alert
delay time. A PCH 1216 or PCH 1218 device has NO alarm relays to inform the safety system of
the wind turbine (or machine) that the monitored vibration level has exceeded the trigger level.
Only with the PCH 1218 it will be possible to check if a certain bit in PDO4 has been toggled from
“0” to “1”.
An Alert alarm can only be set-up if this alarm is “Enabled”:
After the “Enable” field has been marked the following window becomes visible:
Alert Limit:
The alert Limit is the trigger level at which the Alert Alarm indicator in the main Display window
in the PCH Vibration Studio software for that particular condition Band will turn yellow. By default
the Alert Limit is set to 50 % of the full scale.
Alert Delay:
To avoid false alarms because of short pulse like increases in the monitored vibration level it is
recommended to implement a certain delay (minimum) time during which the vibration level must
continuously exceed the Alert alarm trigger level.
By default the Alert Delay time is set to 0 seconds.
Page 42
Alert Latched:
Un-latched alarms will turn back to stand-by as soon as the hang time (in seconds), which is the
minimum time an alarm will remain active after the conditions that triggered the alarm, have ended.
Latched alarms will remain active even if after the conditions that triggered the alarm, have ended.
Very serious alarms are usually latched to make sure the operator notices the alarm.
Danger Alarm Setup:
The set-up of a Danger alarm is very much similar to that of an Alert alarm.
There are only a few differences that should be kept in mind:
- The Danger Limit MUST ALWAYS BE AT A HIGHER LEVEL THAN the Alert Limit
- By default the Danger Limit is set to 80 % of the full scale.
- By default the Danger Delay time is set to 5 seconds. This way the Danger alarm will trigger
faster whenever a (sudden) dangerous vibration level is detected.
!
Note!
Only when specified explicitly in the Customer Specification Agreement the Alert and
Danger alarms in the conditioning bands will be enabled.
8.5 Re-set of Alarms
How alarms are re-set depends on whether these alarms are latched or non-latched.
Non-latched alarms:
An active non-latched Alert or Danger alarm will automatically be cleared when the vibration level
is reduced below its trigger level for a period longer than the hang time of typical 1 second
:
Latched alarms can only be re-set by clicking on the reset alarm button within the PCH Vibration
Studio software or by carrying out a successful self-test
SSD alarms:
SSD alarms are always latched and very “sticky”. An SSD alarm can – according to the GL2010
guide lines – only be reset on-site by a user/operator carrying out a successful SSD test. Off the
record the SSD alarm in a PCH 1218 device can be reset via the CANopen protocol.
Page 43
9 SSD function
A number of wind turbines have been totally damaged by e.g. a failure of one or more of the main
ball bearings in the drive chain of the wind turbine. The very high costs following such a
catastrophic event has made the insurance companies force the wind turbine owners to install a
device that can detect the tremendous shocks that appear during the initial state of the disaster.
A PCH 1216 or PCH 1218 device is provided with an embedded Shock Safety Detection (SSD)
function in order to detect those huge unexpected shocks imposed on the structure of the wind
turbine tower. Other causes of huge vibrations/shocks might be damaged blades, loose main
bearing, loose gearbox or other major events, any of which can create large shock transients that are
able to severely damage the wind turbine on short notice.
9.1 SSD function
The SSD function is a limited configurable function inside a PCH 1216 or PCH 1218 device.
The SSD function is a factory installed function that canNOT be disabled by the user.
The SSD function is certified by Germanischer Lloyd according to their latest guide lines and will
be updated whenever newer versions become available. Whenever the SSD function is included in
the purchase order, the presence of the SSD function is clearly marked on the outside of the monitor
with the following label:
9.2 SSD function configuration
The SSD function will always be pre-configured prior to delivery by the manufacturer PCH
Engineering A/S.
When the SSD function is pre-configured by the factory, the firmware (software embedded in the
vibration monitor) locks the SSD parameters, so they can no longer be altered by the user.
The SSD function is based on two dedicated mechanical relays that will – if connected in series –
form a fully redundant ISO 13849-1:2006 category 3 compliant sub-system. An integrated
diagnostic system monitors the health of the two relays thus ensuring an extremely high level of
functional safety.
During “stand by” the relays are in “make mode” and the contacts at pin #2 and pin #3 are at GND
level, i.e. connected to GND.
When activated the relays toggle to “break mode” and the connection of pin #2 and pin #3 to GND
is interrupted.
Page 44
Those PCH 1216 and PVCH 1218 devices that are provided with the “Two String SSD relay
Function” do possess a second pair of dedicated mechanical SSD relays, which behave exactly the
same as the first pair described above.
!
Warning!
The relays are small highly reliable mechanical relays. Excessive current (and/or
voltage) will destroy a relay. For this reason please do not apply any power source
directly to the relay but protect the relay with a resistor in series that will limit the current.
Please make sure that the following ratings/conditions for the SSD Danger alarm relays
are not exceeded in your application:
- Maximum load voltage
: 48 VDC
- Maximum load current
: 100 mA
- Maximum power
: 5 Watts
The SSD function is pre-configured to work in TriaxDir mode and can by no means be disabled by
the user or operator. Once pre-set, this setting can only be changed by the manufacturer PCH
Engineering A/S. However, it is possible – on user/customer request – to make the SSD function
operable with any one of the three(3) available OmniDir modes. The SSD function will NOT with
one of the SingleDir modes.
All parameters related to the SSD function are factory-set except Full Scale in m/s2 and the SSD
danger alarm trigger level. This can be set by an operator using the PCH Vibration Studio software
to a maximum full scale of 50 m/s2 and a maximum trigger level not exceeding the full scale level.
The filter range is pre-set from 0.1 Hz (Bypass) to 25 Hz (Butterworth Low pass filter) and the
detector type is: Peak. The detector attack time is 5 ms and the decay time is 2 s.
Although the SSD alarm delay time is set to 0 seconds, the time lap between the moment the SSD
Danger alarm trigger level is exceeded until the moment the alarm actually triggers is
approximately 50 ms. The total of 50 ms is within the requirements (<100 ms) by Germanischer
Lloyd.
How to change the SSD Danger alarm parameters is very similar to changing the setup parameters
in a conditioning band. See section 8.4 of this manual for more information on this topic.
9.3 SSD test and re-set of a SSD Danger alarm
According to Germanischer Lloyd a triggered SSD Danger alarm can and must only be reset
manually by a person physically going to the wind turbine nacelle by pressing a SSD test/re-set
button positioned close to the PCH 1216 or PCH 1218 device or on the control panel in the nacelle.
Such a button will connect pin #4 with GND level, which is available at pin #7 for a PCH 1216 or
PCH 1218 device with an 8-pin connector or at pin #12 for a PCH 1216 or PCH 1218 device with a
12-pin connector. Therefore there is no possibility to reset the SSD alarm (remotely) from the PCH
Vibration Studio control software.
Whenever the “SSD test” is activated all processes/components/parts in the SSD measurement
chain are tested, i.e. both accelerometers, conditioning circuit, signal processing and SSD relay
outputs. Activating the “SSD test” will force the SSD relay outputs at pin #2 and pin #3 to become
active – thus performing a simulation of a SSD alarm. Similar simultaneous behaviour will occur at
pin #10 and pin #11 of PCH 1216 and PCH 1218 devices with an embedded two string SSD
function.
Page 45
See sections 4.1 and 4.2 for the layout of the connectors that are relevant for your PCH 1216 or
PCH 1218 device.
If any erroneous behaviour of the SSD function that is detected during a “SSD test” will cause a
system error visible indicated by the device LED on the enclosure by showing a continuous red
light.
If the SSD test ends flawless any activated SSD Danger alarm will be de-activated and the
PCH1216/1218 will return to standby mode, which will be indicated by the same device LED on
the enclosure by showing a continuous green light.
An SSD alarm event is saved in non-volatile memory and will therefore be persistent (i.e. sticky)
across power off/on. This means that the monitor will not be reset or in any other way lose the SSD
alarm status or settings in connection with an (un-)intentional power down. Also a loss of
communication will NOT result in a reset of a triggered SSD Danger alarm.
!
Warning!
The SSD test/re-set is to be carried out by a technician that is personally present in the
wind turbine (nacelle). Since a remote re-set of a SSD Danger alarm is a clear violation
of the Germanische Lloyd GL2010 approval, we strongly advice the users of PCH 1218
devices NOT to use the CANopen protocol to re-set an activated SSD Danger alarm.
Doing so might have serious consequences concerning compensation for damages by
your insurance company. However, if the wind turbine is densely surveyed by means of
cameras (and microphones) that are positioned both inside and outside the wind turbine
thus giving a technician in a remote control room a full overview of the wind turbine’s
operational condition, it is officially allowed to initiate the SSD test from this remote
control room. See the GL2010 certification guide lines, chapter 2.2.2.5 on the subject of
camera surveillance.
In no way can the manufacturer PCH Engineering be made responsible for any damage(s)
or increase of damage(s) caused by a remote re-set of an activated SSD Danger alarm.
Page 46
10 Service and maintenance
10.1 Periodic self-test of a PCH 1216 or PCH 1218 device
As described in chapter 11 of the “Guidelines for Certification of Wind turbines, Edition 2010”
published by Germanische Lloyd (GL) in Germany, every wind turbine is subject to Periodic
Monitoring, which is an inspection of the wind turbine by a technical expert at regular intervals.
The inspection shall be carried out according to the conditions in the Certification Reports.
Inspection intervals are laid down in the corresponding Certification Reports, their annexes, or
indirectly in the form of references. The objective of Periodic Monitoring is the examination
(inspection) of the entire wind turbine including the machinery, the safety devices and the structural
integrity. The typical inspection interval for a modern wind turbine is: once every 6 months.
The PCH 1216 and/or PCH 1218 devices are regarded as a safety device in the wind turbine. To
assure a high level of reliability of the PCH 1216 and PCH 1218 device’s safety functions the
device has a built-in watch dog, which constantly monitors possible faults in the circuitry. To
maximize the integrity of the device even further the manufacturer suggests that the vibration
monitor is tested at least at fixed intervals that correspond with the periodic monitoring procedure,
so that the tests are not in conflict with the normal operation of the wind turbine.
A PCH 1216 or PCH 1218 device has two individual “forced” self-test functions, which can be
triggered independently of each other by the technical expert:
- The first forced self-test function is a general self-test function that tests the complete signal
chain for all functions (excluding the SSD Danger alarm function) from the sensors onwards,
including the fieldbus (if implemented and active).
! Note:
It is highly recommended that this self-test function is initiated during start up and at
regular intervals (6 months, see above) under the condition that the actual vibration
level is below a certain low level, i.e. at low wind speed condition only.
For more information on the SSD test function, see chapter 4.9 of this manual.
- The second forced self-test is a special SSD test, which tests the specific signal chain related to
the “Safety Shock Detector” function.
For more information on the SSD test function, see chapter 4.5 of this manual.
!
Warning!
The SSD test/re-set is to be carried out by a technician that is personally present in the
wind turbine (nacelle). Since a remote re-set of a SSD Danger alarm is a clear violation
of the Germanische Lloyd GL2010 approval, we strongly advice the users of PCH 1218
devices NOT to use the CANopen protocol to re-set an activated SSD Danger alarm.
Doing so might have serious consequences concerning compensation for damages by
your insurance company. However, if the wind turbine is densely surveyed by means of
cameras (and microphones) that are positioned both inside and outside the wind turbine
thus giving a technician in a remote control room a full overview of the wind turbine’s
operational condition, it is officially allowed to initiate the SSD test from this remote
control room. See the GL2010 certification guide lines, chapter 2.2.2.5 on the subject of
camera surveillance.
Page 47
10.2 Servicing a PCH 1216 or PCH 1218
System Failure text
Possible reason
Correction
- Transducer error
A transducer hardware error is
detected during self-test.
Transducer error can only arise
when running the Self-test.
The Monitor needs to be returned to
PCH Engineering for correction of
error. If self-test was performed during
machine running with high level of
vibration the test can come out with an
error. If so run the test again without
machine running
- CRC error in factory settings
(Fatal)
A power surge or power drop out has
corrupted the storage of calibrating
parameters.
The Monitor needs to be returned to
PCH Engineering for correction.
- CRC error in Auto calibration
Not implemented yet
Run self-test. If persistent this due to
malfunction in the transducer(s) or
hardware
- CRC error in setup area
Power surge or power drop out while
writing to the setup.
Reload monitor settings
- Firmware incompatibility
detected in Monitor
Attempt to update monitor with
incompatible firmware. In this state
the monitor report system error and
measuring and monitoring is
disabled.
Update monitor with compatible
Administration Processor firmware
- Monitor is in update mode
The monitor is in firmware update
mode because a firmware update
session was interrupted or attempt to
update with incompatible firmware.
Update monitor with compatible
Administration Processor firmware
- CRC error in vibration setup
Power surge or power drop out while
writing to the setup.
Reload monitor settings
- CRC error in process setup
Power surge or power drop out while
writing to the setup.
Reload monitor settings
- CRC error in Hardware setup
Power surge or power drop out while
writing to the setup.
Reload monitor settings
- Temperature limit exceeded.
The Monitor is installed in a place
that brings Operating Temperature
out of specified range.
Relocate the monitor to a place with
higher or lower temperature.
Inspection of the installation will tell
which limit is exceeded.
- DSP Halted
Turn power of the monitor off and then
on again.
- Scalar output overload band
1 to 4 and/or SSD band 5
DC output is greater than 125 % of
Full Scale Level.
Increase full scale level, lower the
input signal or relocate the monitor to
another vibration pickup point
- Transducer A, B, C input
overload
Input signal amplitude is too large for
the input amplifier to handle. Very
High vibrations e.g. at High
frequencies
Lower input signal or relocate the
monitor to another vibration pickup
point
System Failures Reported by Monitor Status
Page 48
10.3 Maintenance and repair
A PCH 1216 or PCH 1218 vibration monitor device is maintenance free.
! Note:
To clean a PCH 1216 or PCH 1218 device, please wipe its enclosure, connector(s) and
cable(s) with a moisture cloth. If necessary, add a small amount of a mild detergent.
In the unlikely event that you experience erroneous behaviour of the device, e.g. a system failure,
we advise you to carry out a complete self-test of the device for a couple of minutes.
In case the system failure is persistent after such a complete self-test we strongly recommend you to
contact the manufacturer PCH Engineering A/S or its local distributor for the immediate repair of
the device.
Please contact the manufacturer for a Return Merchandise Authorization (RMA) in order to arrange the
return of goods to the supplier and to have the product repaired or replaced or in order to receive a
refund or credit for another product within the product's warranty period.
Whenever you contact the manufacturer for a repair of the device you are kindly requested to have the
following information at hand:
- The type number of the vibration monitor: PCH 1216 or PCH 1218
- The CHF 836xxx(Ry) or CHF 838xxx(Ry) number of the device (where “xxx” is a 3-digit number
and “y” is single-digit number)
- The serial number of the device, visible on a label on the enclosure
- The colour and behavior (continuous or blinking with 1 s intervals) of the status LED on the
enclosure
- The status indicators in the PCH Vibration Studio software that are active, if any
Manufacturer’s shipping address:
PCH Engineering A/S Phone : +45 4576 8776 Email : pch@pch-engineering.dk
Ved Klaedebo 4 Fax : +45 4576 8702 Internet : http://www.pch-engineering.dk
DK-2970 Hoersholm
Denmark
Page 49
11 Installation
11.1 Delivery of standard PCH 1216 or PCH 1218 device
When you take delivery of one or more standard PCH 1216 or PCH 1218 devices you should
receive the following items:
- One or more PCH 1216 or PCH 1218 device main unit(s) completely assembled and configured
to your version number CHF836xxxRy or CHF838xxxRy respectively as mentioned in your
Customer Specification Agreement, if any
- Upon request a free of charge paper printed or CD-ROM version of this user manual
- Upon request a free of charge copy of the CHT 1044 PCH Vibration Study control software on
CD-ROM
- an optional CHB 1131P8 (8-pin connector) or CHB 1131P12 (12-pin connector) Service Cable
- an optional CHB 1134P8 (8-pin connector) or CHB 1134P12 (12-pin connector) Service Adaptor
11.2 Installation in general
The PCH 1216 or PCH 1218 device you got delivered should be rigidly mounted inside the nacelle,
e.g. directly to the main frame or similar. A shielded data cable with a matching A-coded 5-pin, 8-
pin or 12-pin female screw plug must be used to assure an ingress protection level of up to
IP66/IP68.
There are several manufacturers of suitable 5-pin and 8-pin plugs like ERNI and BINDER. To find
a suitable 12-pin plug is much more of a challenge. The manufacturer PCH Engineering A/S
recommends the following 12-pin plug:
- Hirose (JP) type LF10WBP-12S
The shielding of the cable must be connected to the monitor housing and in the opposite end to the
Topbox or PLC/Controller housing/enclosure. This ensures an optimal Faraday cage for the best
possible EMI/RFI protection.
Please make sure that the locks of any connectors are firmly tightened.
Page 50
Delivery of PCH 1216 or PCH 1218 devices for OEM:
In order to comply with specific customer requirements the PCH 1216 or PCH 1218 devices can be
delivered in a large number of varieties. Please contact PCH Engineering for more information on
special solutions and OEM deliveries.
! Warning:
The correct functioning of a PCH 1216 or PCH 1218 device can only be guaranteed by the
manufacturer if any accessories that are used together with the device are either designed
for or approved/certified by the manufacturer PCH Engineering A/S (Denmark).
! Warning:
A PCH 1216 or PCH 1218 device has no internal forced ventilation. These devices have no
external ventilation requirements as long as the surface temperature during operation does
not exceed the maximum ambient temperature value Ta stated in section 12 of this manual:
“Technical Data”.
本站仅提供存储服务,所有内容均由用户发布,如发现有害或侵权内容,请点击举报
打开APP,阅读全文并永久保存 查看更多类似文章
猜你喜欢
类似文章
【热】打开小程序,算一算2024你的财运
``XP系统全面优化``
Gartner Reprint
General Network Engineer Interview Questions ...
Mobile Phone As Home Computer
Internet - MSN Encarta
C#的发展历程及应用范围
更多类似文章 >>
生活服务
热点新闻
分享 收藏 导长图 关注 下载文章
绑定账号成功
后续可登录账号畅享VIP特权!
如果VIP功能使用有故障,
可点击这里联系客服!

联系客服